Welcome To BTX Online !   

Enhancer 3000 Monitoring Systems

More Views


Enhancer 3000 Monitoring Systems
Monitor critical parameters while performing electroporation applications


Product Item # Price  
Enhancer 3000 with Interface Box, Probe and Cables (N. America) 45-0057   View Price
Enhancer 3000SC with Interface Box, Oscilloscope, Communications Module and Cables (N. America) 45-0059   View Price
Legacy ECM 630 Electroporation System Complete, for Cuvettes with Monitoring 45-0071   View Price
Legacy ECM 630 High Throughput Electroporation System with Monitoring 45-0471   View Price
Legacy ECM 830 High Throughput Electroporation System with Monitoring 45-0472   View Price
Legacy ECM 2001 Electroporation System Complete, for Cuvettes with Monitoring (N. America) 45-0013   View Price
Legacy ECM 2001 Electroporation System Complete, for Cuvettes with Monitoring 45-0013INT   View Price
Enhancer 3000 High Voltage Interface Box 45-0062   View Price
Enhancer 3000 High Voltage Probe 45-0063   View Price
Enhancer 3000 Oscilloscope, Tektronix TDS1002B w/o 1X/10X Low Voltage Probes, 2-Channel, 60 mHz 45-0064   View Price
Login/Register to shop in US only.
All others please Login/Register or contact us for local pricing/quotes.

 

 

Details

The Enhancer 3000 Monitoring System allows researchers to maintain efficiencies, optimize both electrical and biological parameters, view sample runs, troubleshoot possible problems and easily perform routine quality control. Communications using the USB data port permits storage and documentation of data for further analysis.

The Enhancer 3000 simply interfaces between the generator and the electroporation chamber (safety stand or BTX electroporation plate handler). In addition, the optional communications module allows the user to send the waveform information directly to a printer or a computer.

Key Features
  • Confirms and tracks all key electroporation parameters including waveform, pulse lengths, peak amplitude, pulse intervals, field strength, and AC duration
  • Monitors all waveforms for optimization or troubleshooting of electroporation parameters
  • Facilitates calculation of field strength, pulse length, pulse duration or frequency
  • Safe connections protect user from exposure to high voltage
  • Optional Communications Module offers standard printer and computer interface

 Included Items

Item # Name Included Items
45-0057 Enhancer 3000 Interface box and cables
45-0059 Enhancer 3000SC Interface box, Oscilloscope, communications module and cables
45-0071 ECM 630 Enhancer System ECM 630 Exponential Decay Wave Electroporation System (45-0001) and Enhancer 3000 (45-0059), Enhancer 3000 Probe, Enhancer Interface Box, Oscilloscope, Communications Module
45-0072 ECM 830 Enhancer System ECM 830 Square Wave Electroporation System (45-0002) and Enhancer 3000 (45-0059), Enhancer 3000 Probe, Enhancer Interface Box, Oscilloscope, Communications Module
45-0013 ECM 2001 Enhancer System ECM 2001 Electroporation System (45-0011) and Enhancer 3000 (45-0059), Enhancer 3000 Probe, Enhancer Interface Box, Oscilloscope with USB communications
45-0471 HT 630 Enhancer System ECM 630 Generator (45-0051), 2 mm HT Plates, HT 200 Plate Handler (45-0401) and Enhancer 3000 (45-0059)
45-0472 HT 830 Enhancer System ECM 830 Generator (45-0052), 4 mm HT Plates, HT 200 Plate Handler (45-0401) and Enhancer 3000 (45-0059)

   Notes: Communications module option is required to support printing and computer connection. Digital storage oscilloscope is required to display waveforms.

 

Enhancer 3000 High Voltage Probe Specifications

Attenuation 1:100 / 1:1000
Bandwidth DC to 70 MHz (-3 dB)
Accuracy +/-2%
Input Impedance 10 M Ω// 10 pF each side to ground
Input Voltage:  
   Differential Range +/-700 V (DC+Peak AC) or 700 Vrms @ 1/100
+/-7000 V (DC+Peak AC) or 5000 Vrms @ 1/1000
   Common Mode Range +/-7000 V (DC+Peak AC) or
2500 Vrms @ 1/100 & 1/1000
   Absolute Max. Range

1 ±7000 V (DC+Peak AC) or 2500 Vrms
@ 1/100 & 1/1000 in common mode
±7000 V (DC+Peak AC) or 5000 Vrms
@ 1/100 & 1/1000 in differential mode

Output Voltage:  
   Max. Amplitude +/-7 V (into 50k ohm load)
   Offset (typical) < +/-5 mV
   Noise (typical) 0.9 mVrms
Source Impedance (typical) 50 Ω
CMRR (typical) -80 dB @ 60 Hz, -50 dB @ 1 MHz
Operating Temperature -10 to 40°C
Ambient Storage Temperature  -30° to 70°C
Operating Humidity 25 to 85% RH
Ambient Storage Humidity 25 to 85% RH
Pollution Degree 2 (Not to be operated in conductive pollutants atmosphere
Power Requirements:  
   Standard 4 AA cells or optional mains adapter or regulated 9 VDC/120mA mains adapter
Length of BNC Cable 88.9 cm (35 in)
Length of Input  Leads 30.48 cm (12 in)
Weight 454 g (1 lb)
Dimensions (L x W x H) 20.3 x 8.3 x 3.8 cm (8" x 3.25 x 1.5 in)

 

Notes: 1. Voltage limit isthe lesser of the DC + Peak AC and RMS values. 2. a. The supplied voltage must be lessthan 12 V and greater than 4.4 V, otherwise the probe
could be damaged or will not operate properly. b. Polarityis“+”inside and “-” outside.For wrong polarity, a built-in circuit protectsthe probe; no danger or damage will occur.c. When the voltage of the batteries become too low, the power indicator will flicker.

 

Voltage Interface Box Specifications

Input, Output, Probe Connections:   
   Type 4 mm Banana Jack
   Voltage 2500 Vrms, 3500 Vpk CAT II
   Current (Max. Continuous) 15 A
   Current (Peak) 1000 A for 100 µs
Dimensions (L x W x H) 13.5 cm x 13.5 cm x 5.8 cm (5.3 x 5.3 x 2.3 in) 

 

High Voltage Cable Specifications

Type Retractable Sheath Banana Plug
Length 91.44 cm (3 ft)
Wire Gauge 18 AWG
Insulation PVC
Voltage Rating 1000 Vrms IEC 1010-2-31, Category III
Current 10 A
Operating Temperature +55º C (+131º F) Max.
  • Optimize and troubleshoot electroporation settings
  • Capture and print results for documentation purposes
  • Track and download images to computer for analysis
  • Compatible with the ECM 630, ECM 830 and ECM 2001