Petri Dish Electrode


Petri Dish Electrode
To electroporate adherent cells or tissue grown in a petri dish that functions as the electroporation chamber
Product Item # Price  
Petri Dish Electrode 45-0100   View Price
Login/Register to shop in US only.
All others please Login/Register or contact us for local pricing/quotes.

 

 

Details

The Petri Dish Electrode is designed to be used with a 100 mm Petri dish that functions as the electroporation chamber. The Petri Dish Electrode is used to electroporate adherent cells or tissue grown in a petri dish.

To perform electroporation, simply add the exogenous molecule of interest into the electroporation buffer. The buffer can range in volume from 10 ml to 50 ml and is added to the cells grown in the plate. The electrode is lowered into the petri dish containing the sample and pulsed.

The electrode assembly has a 2 mm gap size. It contains parallel stainless steel electrodes which generate a homogeneous field. The Petri Dish Electrode is compatible with Gemini X2, ECM 630, ECM 830, ECM 2001+, and Legacy ECM 2001 Generators. The Petri Dish Electrode connects directly to Gemini X2, ECM 630, and ECM 830 generators. Please see table and connection diagram below for required cables and adapters for use with ECM 2001+ and Legacy ECM 2001.

REQUIRED CABLES

Required for connection to ECM 2001+
45-0088 Female/Female Adapter Set for Banana Plug Cables
45-2057 or 45-2058 ECM 2001+ HV Output Cable, 5 ft or 10 ft length
Required for connection to Legacy ECM 2001
45-0088 Female/Female Adapter Set for Banana Plug Cables
45-0083 Coaxial to Banana Plug Cables, Red and Black, 10 ft

 

USE OF PETRI DISH ELECTRODES WITH THE GEMINI X2 AND ECM 830 ELECTROPORATION SYSTEMS

Gemini X2 and ECM830 generators connect directly to the Petri Dish Electrode. 

USE OF PETRI DISH ELECTRODES WITH THE ECM 2001+ ELECTROFUSION & ELECTROPORATION SYSTEM

In this example, the ECM 2001+ generator is connected using the sequence of cables and adapters illustrated below, then connected to the Petri Dish Electrode. 

+ +
 ECM 2001+ System  

 45-2057 or 45-2058 HV Output Cable

   45-0088 Female/Female Adapters 
Petri Dish Electrode

 

USE OF THE PETRI DISH ELECTRODES WITH THE LEGACY ECM 2001 ELECTRO CELL FUSION & ELECTROPORATION SYSTEM

In this example, the ECM 2001 generator is connected to the electrodes using the sequence of cables and adapters illustrated below, then connected to the electrode. This electrode is also compatible with the ECM 630, ECM 830, ECM 2001+ and Gemini X2 systems. 

+ +
 Legacy ECM 2001 System    45-0083 Coaxial Cables    45-0088 Female/Female Adapters 
Petri Dish Electrode

Generator Compatibility ECM 630, ECM 830, ECM 2001+, Gemini X2, and Legacy ECM 2001
Voltage Range 0 to 2000 V 
Volume Range 10 to 50 ml 
Gap Size 2 mm
Autoclavable No
Field Type Homogeneous
Pulse Length 10 µsec to 10 msec 

  • Adherent mammalian cell transfections
  • Plant tissue cell transfections