Loading... Please wait...The Enhancer 3000 Monitoring System allows the researcher to confirm and track key electroporation parameters, including waveform, peak voltage, field strength, pulse length, pulse duration, and frequency in all electroporation generators with external pulse capabilities.
These parameters are critical during optimization sample runs or troubleshooting.
The Enhancer 3000 simply interfaces between the generator and the electroporation chamber (safety stand or BTX electroporation plate handler),
In addition, the optional communications module allows the user to send the waveform information directly to a printer or a computer.
Key Features
|
Enhancer 3000 High Voltage Probe Specifications: |
||
|
Attenuation: |
|
1:100 / 1:1000 |
|
Bandwidth: |
|
DC to 70MHz (-3dB) |
|
Accuracy: |
|
+/- 2% |
|
Input Voltage: |
|
|
|
-Differential Range: |
|
+/- 700V(DC+Peak AC) or 700Vrms @ 1/100 |
|
-Common Mode Range: |
|
+/- 7000V(DC+Peak AC) or 2500Vrms @ 1/100 & 1/1000 |
|
Output Voltage: |
|
|
|
-Max. Amplitude: |
|
+/- 7V (into 50k ohm load) |
|
-Offset (typical): |
|
< +/- 5mV |
|
Ambient Operating Temperature: |
|
-10 to 40°C |
|
Ambient Operating Humidity: |
|
25 to 85% RH |
|
Power Requirements: |
|
|
|
-Standard: |
|
4 AA cells or optional mains adapter |
|
Length of BNC Cable: |
|
35" |
|
Length of Input Leads: |
|
12" |
|
Weight: |
|
1 lb |
|
Dimensions (LxWxH): |
|
8" x 3 ¼" x 1 ½" |
|
Enhancer 3000I High Voltage Interface Box Specifications: |
||
|
Input, Output, Probe Connections: |
||
|
Type: |
|
4mm Banana Jack |
|
Voltage: |
|
2500Vrms, 3500Vpk CAT II |
|
Current (Max. Continuous): |
|
15A |
|
Current (Peak): |
|
1000A for 100us |
|
Dimensions (LxWxH): |
|
5.3" x 5.3" x 2.3" (13.5cm x 13.5cm x 5.8cm)
|
|
45-0057 |
Enhancer 3000 |
Enhancer 3000, Includes Interface box and cables |
|
45-0059 |
Enhancer 3000SC |
Enhancer 3000, includes interface box, Oscilloscope, communications module and cables |
|
45-0071 |
ECM 630 Enhancer |
ECM 630 Enhancer SYSTEM, Includes 45-0001 and 45-0059 |
|
45-0072 |
ECM 830 Enhancer |
ECM 830 Enhancer SYSTEM, Includes 45-0002 and 45-0059 |
|
45-0013 |
ECM 2001 Enhancer |
ECM 2001 Enhancer SYSTEM, Includes 45-0011 and 45-0059 |
|
45-0471 |
HT 630 Enhancer |
HT 630 Enhancer SYSTEM, Includes 45-0051, 2 mm HT Plates, 45-0401 and 45-0059 |
|
45-0472 |
HT 830 Enhancer |
HT 830 Enhancer SYSTEM, Includes 45-0052, 4 mm HT Plates, 45-0401 and 45-0059 |
BTX Monthly Tech Trends Articles
For further references regarding specific applications and optimization, please contact BTX Technical Support:
BTX-Division of Harvard Apparatus
84 October Hill Road
Holliston, MA 01746
Phone: 1-508-893-8999
Toll Free: 1-800-272-2775
Fax: 1-508-429-5732
Email: techsupport.btx@harvardapparatus.com
If outside the United States or Canada: call 508-893-8999 or
contact your nearest BTX Distributor